Calibration & verificationAll calibration servicesPipette calibrationCentrifuge / RPM verificationThermometer calibrationPCR thermocycler calibrationCO₂ incubator calibrationElectrical safety (NFPA 99)EKG / ECG certificationFume hood certificationBiosafety cabinet certificationClean room & USP 797 certificationEquipment service & repairLab equipment repairHistology & pathology serviceCryostat repair & PMScientific refrigerationEngineering & modernizationEmergency service & loanersRental & loaner instrumentsDecontamination & removalModels we serviceCompliance & lab setupInspection readinessService agreementsMohs lab equipmentCryostat Care PlanMohs staffing & risk managementLab setup & turnkey labsClient portal (coming soon)Free readiness checklistMedical offices & urgent careDental practicesGI & endoscopy centers Who we serveClinical laboratoriesMolecular & research labsPathology & Mohs labsIVF & fertility labsDermatology groupsMedical-center groupsCommunity health centersResidential & specialty careSee all facilities we serve Where we workMiami & Miami-DadeFort LauderdaleWest Palm BeachTampa BayOrlandoNaples & Fort MyersPuerto Rico MoreResourcesReviews Book a free assessment
Lab Safety Guide

Fume hood certification: what's tested and how often

BMCA Group · · 3 min read · Florida & Puerto Rico

A chemical fume hood is the main thing standing between your staff and solvent vapors, formalin and xylene. It only works if it is moving enough air, in the right pattern, at the sash height people actually use. Certification is how you prove it does, and it is one of the first safety records an inspector asks for in a pathology or histology lab.

How often fume hoods must be certified

Fume hoods are typically certified at least once a year. The ANSI/ASSP Z9.5 laboratory ventilation standard calls for routine performance testing, and most safety programs and accrediting bodies treat annual as the minimum. A hood should also be tested:

  • When it is first installed, before use
  • After it is moved, even within the same room
  • After changes to the building exhaust or HVAC system
  • After repairs to the hood, sash or airflow monitor
  • Whenever users suspect a problem, such as odors at the sash

What the test measures

Face velocity

The certifier measures air speed across a grid of points over the sash opening at the working sash height, then averages it. Most conventional hoods are held to an average of roughly 80 to 120 feet per minute, though your facility or the manufacturer may set a different target. Individual low readings matter as much as the average.

Airflow visualization

Smoke is released around the sash edges, work surface and inside the hood to show that air flows in and stays in. Smoke rolling back out at the sash is a containment failure even if the face velocity looks fine.

Sash and alarm

The sash should move smoothly and hold its position, and the airflow monitor or low-flow alarm should trip when airflow drops. A muted or disabled alarm is one of the most common findings.

Containment testing

For new installations, or where your safety program requires it, testing to ASHRAE 110 adds a tracer gas test that measures how much escapes the hood under controlled conditions.

What you should receive

  • A report with every face-velocity reading, the average and the target
  • Sash height at which the hood was tested
  • Smoke test and alarm results
  • Date, certifier and next-due date
  • A label on the hood that matches the report, showing the maximum safe sash height

Fume hood or biosafety cabinet?

They are not interchangeable. A chemical fume hood protects the user from chemical vapors and exhausts air out of the building. A biosafety cabinet uses HEPA filters to protect the user, the sample and the room from biological material. They are certified to different standards: fume hoods to Z9.5 and ASHRAE 110, biosafety cabinets to NSF/ANSI 49. See biosafety cabinet certification for the cabinet side.

BMCA schedules and documents fume hood certification across Florida and Puerto Rico, usually on the same visit as your biosafety cabinets and calibrations. The field testing is performed by our certification partner, Medrep Technologies.

Common questions

How often does a fume hood need to be certified?
At least annually, and also at installation, after relocation, after HVAC or exhaust changes and after repairs.
What face velocity should a fume hood have?
Most conventional fume hoods are held to an average face velocity of about 80 to 120 feet per minute at the working sash height. Your facility or the manufacturer may set a different target.
Is a fume hood the same as a biosafety cabinet?
No. A fume hood protects the user from chemical vapors. A biosafety cabinet filters air through HEPA filters to protect the user, the sample and the room from biological material.

See where your lab stands — free.

Request a no-obligation Compliance & Equipment Readiness Assessment. We'll spot-check your equipment and documentation and leave you a short readiness report.

No obligation · we'll reply within one business day.
Prefer to talk? Call 305.238.5326
Free readiness assessment

Not sure where your lab actually stands?

We'll spot-check your equipment and documentation against what CLIA, CAP, and AHCA surveyors check, then leave you a short readiness report — no obligation, whether or not you ever work with us.

☎ 305.238.5326Free assessment